Composition :

O-Phthalaidehyde: 0.55% w/w with Chelating Agents, Corrosion Inhibitors, and other inert ingredients.


Direction of use :

1. Device Cleaning / Decontaminations

Blood. other body fluids. and lubricants must be thoroughly cleaned from the surface and lumen of medical devices before reprocessing In the disinfectant. Before immersion in Germiklean OPA solution, thoroughly clean device including all lumen, using a cleaning protocol or standard.

2. Solution Usage

No activation is required. Record date of open on the container label, or in a log book. After opening, the solution remaining in the container can be used for up to 75 days. Record the date the solution was poured out of the original container into a secondary container in a log book. or on a label affixed to the secondary container that can be used for a period up to 14 days and to be discarded after that.

3. High Level Disinfection

  • Manual Processing: Immerse device completely in Gerrniklean OPA solution for a minimum of 12 minutes at 20oC or higher to destroy all pathogenic microorganisms.
  • Automatic Endoscope Reprocessor to be set to a minimum of 25oC with minimum 5 minutes of immersion time. Following removal from Germiklean OPA solution, thoroughly rinse the medical device by immersing in sterile water. Keep device totally immersed for a minute. Repeat the procedure twice.

Precautions :

  • Do not get it in eyes, skin or clothing
  • Flush Eyes with cold water In case of contact with the eyes
  • Wear personal protective gear before handling
  • Wash thoroughly with soap and water after handlIng